Hemoglobinopathies

The principle of the Targeted MS/MS Hemo screening assay is based on a rapid hemoglobin (Hb) extraction and denaturation from patients’ DBS samples before its digestion by trypsin. The digested extract is then analysed by MS/MS, in flow injection analysis mode (FIA), on a “triple quadrupole” system. The MS/MS mode used for this assay is the Multiple Reaction Monitoring (MRM): two product ions are selected by peptide and the twelve peptides are simultaneously detected by scanning the two specific “parent ion/product ion couple” for each.

A dedicated software allows time saving in results analysis. Targeted MS/MS Hemo Software (based on Excel) is used to calculate peptide signals ratios, which are analysed by tree-like algorithms. It facilitates the identification of different medically significant phenotypes.

General information

Format: 96 – 480 determinations
Storage: At 2-8°C
Automated Interpretation: Yes
Turn around time: Around 2.5 hours

SKU: HBG-009 Categories: ,
Share :

Description

The newborn screening kit for hemoglobinopathy evaluation, Targeted MS/MS Hemo, is a qualitative analytical test for newborn screening for hemoglobin (Hb)-related conditions such as sickle cell disease. This test is based on the detection by mass spectrometry of peptides generated during the digestion of normal Hb and/or variants (HbA, HbA2, HbC, HbDPunjab/Los Angeles, HbE, HbF, HbOArab and HbS) of the patient.

This assay is dedicated to professional use in diagnostic laboratories. The device is not for self-testing.